MENTOR TEXAS MENTOR MEMORYGEL BOOST BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
SMPB545 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Local Reaction (2035)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 61-year-old caucasian female patient underwent a breast implantation procedure with a mentor memorygel boost breast implant 545cc and experienced swelling/pain and device migration on the left side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent removal and replacement with a similar device on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: local reaction, device migration.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On may 30, 2023, the mentor failure analysis lab received the device for evaluation.Additional information was received with the device indicating the patient identifier is (b)(6).The patient's date of birth was identified as (b)(6) 1960.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the smo ro mod pro boost gel 545cc returned device.Swelling is a temporary normal post-operative condition.Depending on presentation, delayed swelling may require additional work-up by the surgeon.Swelling is a known complication associated with these devices and is referenced in our current product insert data sheet.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.It should be noted that as part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|