ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 6" (15CM) APPX 0.33ML, SMALLBORE BIFUSE EXT SET W/2 MICROCLAVE¿ CLEAR, 3 CLAMPS,; STOPCOCK, I.V. SET
|
Back to Search Results |
|
| Model Number |
011-MC3316 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Codes |
Chemical Exposure (2570); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is available for evaluation, however has not been received.
|
| |
|
Event or Problem Description
|
|
The event involved a 16" (15cm) appx 0.33ml, smallbore bifuse ext set w/2 microclave¿ clear, 3 clamps, rotating luer.It was reported that a disconnection had occurred between the bioconnector and the braun tree, while docetaxel was being administered, generating a cytotoxic spill on the patient's skin.The event occur during patient use, there was a delay in therapy, however no one was harmed as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received on 6/22/2023 for evaluation.One new and one used sample list #011-mc3316, one used b.Braun infusion set, one cloruro sodico 0,9% baxter viaflo 50 ml intravenous bag, one cloruro sodico 0,9% baxter 250 ml intravenous bag with docetaxel, one cloruro sodico 0,9% fresenius kabi freeflex 50 ml intravenous bag and one used perfusion intravenosa 50 ml bottle were returned for evaluation.As received no damage or abnormally were observed in icu medical product.Both sets were leaked tested according procedure and no leaks were observed along the device, male luer were found within specification.Complaint of leaks cannot be confirmed based in the physical sample evaluation (used and new sample).
|
| |
|
Search Alerts/Recalls
|
|
|