| Model Number |
0998-00-3023-83 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to use the cs300 intra-aortic balloon pump (iabp) screen is not working.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information poc: (b)(6).A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.The fse performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released and cleared for clinical service.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|