• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY2 IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY2 IMPLANT; DENTAL IMPLANT Back to Search Results
Catalog Number 834716
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Codes Granuloma (1876); Failure of Implant (1924); Unspecified Infection (1930); Inflammation (1932); Pain (1994); Osteopenia/ Osteoporosis (2651)
Date of Event 10/21/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Per complaint (b)(6), after clinical procedure, patient experienced failure of implant to osseointegrate.
 
Additional Manufacturer Narrative
Patient's identifier, age, sex, weight, other relevant history, including preexisting medical conditions, implant date and explant date were not provided.If the requested information becomes available, a supplementary report will be submitted.The implant failure modes, failure/ loss of osseo-integration and lack of primary stability are not product related and rather are attributed to patient contraindications, conditions, or clinician error in surgical protocol.No product investigation or corrective actions required.Complaints will continue to be trended.
 
Additional Manufacturer Narrative
Follow-up submitted for additional information.Updated section b4 for report submission date, g1 for follow-up report submitter, g3 for awareness date and g6 for report type and follow-up number.Updated h2 for follow-up type.Updated h6 for adverse event problem.Updated section a1, a2, a3, b7, d6a & d6b for patient identifier, age, sex, oral hygiene, other relevant history, implant date and explant date.Patient weight is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGACY2 IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
Manufacturer Contact
sergio alvarado
3050 east hillcrest drive
thousand oaks, CA 91362
8057785975
MDR Report Key17010013
Report Number3001617766-2023-02243
Device Sequence Number6975455
Product Code DZE
UDI-Device Identifier10841307101802
UDI-Public10841307101802
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K192221
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial,Followup
Report Date (Section B) 06/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/10/2024
Device Catalogue Number834716
Device Lot Number156008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/02/2023
Supplement Date Received by Manufacturer06/20/2023
Initial Report FDA Received Date05/26/2023
Supplement Report FDA Received Date06/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-