| Model Number |
FT17LN35NGE059N |
| Medical Device Problem Codes |
Fracture (1260); Inflation Problem (1310); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other text: d4: expiration date and h4: manufacture date are unknown, no lot number has been provided.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
A customer reported that an emergency tracheostomy change had to be performed due to the tracheostomy cuff not inflating sufficiently.It was stated also on the inspection, it appeared that there's a hole/split in the cuff of the tracheostomy.Therefore water leaks out and the cuff does not inflate.No report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).No product or photos were returned therefore no device analysis could be completed.A device history record (dhr) review showed there were no discrepancies or non-conformances during the manufacturing of the reported lot number.If the product is returned the manufacturer will re-open the complaint for further device analysis.
|
| |
|