OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", BIPOLAR, 5 X 330 MM, PREPARATION FORCEPS, MARYLAND, SHORT
|
Back to Search Results |
|
| Model Number |
WA64320C |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During inspection and testing, the teflon body was damaged by burns.And the proximal end was deformed.A review of the device history record found no deviations, that could have caused or contributed to the reported event.Based on the results of the legal manufacturer's investigation, the damaged teflon body was likely, caused by wrong handling by the user.The damage was most likely, caused by a massive high-frequency current flashover.This may be triggered by unintentional contact with other equipment, or by permanent contact between the distal jaws without tissue contact, during a high-frequency application.This leads to a short circuit inside the device.The deformed proximal end was likely, caused by wrong handling.More specifically, subjecting the device to excessive force.The deformation of the proximal end is most likely, attributable to improper combination of the jaws insert and to the use of excessive force by the customer.It was still possible, to combine the jaws insert.However, due to the degree of the deformation, it can no longer be guaranteed that the overload protection mechanism is triggered safely and properly.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
Olympus was informed, that seven (7) hiq+ bipolar maryland graspers broke while performing total laparoscopic hysterectomy (tlh) surgeries.During the past year, all seven (7) are separate events.It was reported, on some of the graspers, sparking was observed, on the tip when using with an electrosurgical generator.No device component or fragment fell inside the patient.There was only a 3¿5 minute delay to replace the device.As the intended procedures were completed with another like model grasper.The device was inspected before use by the nurse, prior to the surgery.No death or injury and no harm or impact to patient or other was reported to olympus.The first event is reported under patient identifier (b)(6), the second event is reported under patient identifier (b)(6), the third event is reported under patient identifier (b)(6) (this report), the fourth event is reported under patient identifier (b)(6), the fifth event is reported under patient identifier (b)(6), the sixth event is reported under patient identifier (b)(6) the seventh event is reported under patient identifier (b)(6), the subject device was sent back to olympus for evaluation.During inspection and testing, the subject device was found to have damaged insulation.This report is being submitted for the malfunction found during evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|