Catalog Number 18965080S
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Migration (4003)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/24/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.H3 other text : device disposition unknown.
 
Event or Problem Description
It was reported that the doctor put a nail in to help a broken tibia and 8-10 days later patient went in and had the staples taken out and was bandaged up.A few hours (3-4) after returning home the patient noticed blood around the bandaged area and the screw backing out.Patient called the doctor and was advised to go the er.Screw was removed when patient went to the er.Screw protruding from left side ankle.Ed recommended wearing a non-weight bearing boot.Pt was provided a prescription for antibiotics.
 
Additional Manufacturer Narrative
The reported event could be confirmed, from the pictures provided for evaluation and matches the alleged failure mode.The device inspection was not possible as the product was not returned for investigation.The device history record could not be reviewed because the affected device was not returned, and the lot number was not communicated.A review of the labeling did not indicate any abnormalities.Few x-rays were presenting no implanted screw and few pictures of partially back out screw and completely removed screw was provided which were presented to our health care professional for evaluation, and a question about the most likely reason for this event, to which the medical expert replied: due to lack of good quality x-rays and with the limited available pictures we cannot say what happened here.There is no picture where the final construct is shown in 2 or more directions.Based on the information provided the exact root cause could not be determined.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
It was reported that the doctor put a nail in to help a broken tibia and 8-10 days later patient went in and had the staples taken out and was bandaged up.A few hours (3-4) after returning home the patient noticed blood around the bandaged area and the screw backing out.Patient called the doctor and was advised to go the er.Screw was removed when patient went to the er.Screw protruding from left side ankle.Ed recommended wearing a non-weight bearing boot.Pt was provided a prescription for antibiotics.
 
Event or Problem Description
It was reported that the doctor put a nail in to help a broken tibia and 8-10 days later patient went in and had the staples taken out and was bandaged up.A few hours (3-4) after returning home the patient noticed blood around the bandaged area and the screw backing out.Patient called the doctor and was advised to go the er.Screw was removed when patient went to the er.Screw protruding from left side ankle.Ed recommended wearing a non-weight bearing boot.Pt was provided a prescription for antibiotics.
 
Additional Manufacturer Narrative
Please note the corrections to d1, d4 catalog #, d4 lot #, h6 device and h6 method codes.The reported event could be confirmed, from the pictures provided for evaluation and matches the alleged failure mode.The device inspection was not possible since the affected device was not returned but pictures were provided which confirms screw back out.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Few x-rays were presenting no implanted screw and few pictures of partially backout screw and completely removed screw was provided which were presented to our health care professional for evaluation, and a question about the most likely reason for this event, to which the medical expert replied: due to lack of good quality x-rays and with the limited available pictures we cannot say what happened here.There is no picture where the final construct is shown in 2 or more directions.Based on the information provided the exact root cause could not be determined.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X80 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17010922
Report Number0009610622-2023-00176
Device Sequence Number12653856
Product Code HSB
UDI-Device Identifier04546540202574
UDI-Public04546540202574
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K203819
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number18965080S
Device Lot NumberK0D3B03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/03/2023
Supplement Date Received by Manufacturer07/06/2023
10/10/2023
Initial Report FDA Received Date05/26/2023
Supplement Report FDA Received Date08/01/2023
11/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexMale
Patient Weight104 KG
-
-