• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ACP KIT SERIES I; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ACP KIT SERIES I; SYRINGE, PISTON Back to Search Results
Model Number ACP KIT SERIES I
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2023
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
On (b)(6) 2023, it was reported by a facility representative via phone that an abs-10011 arthrex acp® kits syringe leaked all the blood out.This occurred during a prp injection on (b)(6) 2023, when they removed the syringe all the blood leaked out of it and another kit was used to complete the injection.
 
Manufacturer Narrative
This report is being submitted to provide additional information in h3, h6: type of investigation, investigation findings, investigation conclusions, and updated information in g3.Since the complaint device was not returned, a physical evaluation of the device was not performed.However, the reported event is not confirmed without the device being returned or any photo provided.The most likely cause can be attributed to user error due to excessive pressure used when drawing back the syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACP KIT SERIES I
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key17011531
MDR Text Key317073985
Report Number1220246-2023-06791
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00888867001824
UDI-Public00888867001824
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK070069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 06/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACP KIT SERIES I
Device Catalogue NumberABS-10011
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/04/2023
Initial Date FDA Received05/26/2023
Supplement Dates Manufacturer Received05/15/2024
Supplement Dates FDA Received06/13/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-