| Model Number |
IPN915189 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use the "applier is misaligned".No patient involvement.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use the "applier is misaligned".No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The complaint sample was not returned.The customer provided photos for investigation.Tecomet, the manufacturing site reported "per customer submitted information the dhr for the alleged defective instrument was reviewed and found complete without any irregularities.This instrument was produced at the tecomet, kenosha, wi facility as part of a (b)(4).Lot in september of 2022.It was found that this order was made from the correct materials and components and all approved processes were followed.This instrument has not been returned for review or evaluation therefore we are unable to determine what may have caused the alleged defect or validate the alleged complaint.All (b)(4) instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed." teleflex will continue to monitor and trend for reports of this nature.
|
| |
|