| Model Number |
212186 |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Inner mechanism loose, oil leaked.Case type / application: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event.An event regarding foreign matter involving a mako saw attachment was reported.The event was not confirmed because the product was not available for inspection.Method & results.-product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there are other complaints with a similar failure mode for this lot.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|
Event or Problem Description
|
|
Inner mechanism loose, oil leaked.Case type / application: tka.
|
| |
|