SMITH & NEPHEW, INC. POROUS TIBIAL TRAY SIZE 2-RIGHT MEDIAL; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, UNICOMPARTMENTAL/UNICONDYLAR, Back to Search Results
Catalog Number 1-10012-250
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10.Smith & nephew, inc., has initiated a field action to voluntarily remove the engage cementless partial knee system from the market.This complaint was originally processed in the engage surgical complaint management system prior to acquisition by smith & nephew, inc.The event has been reassessed and is being retrospectively reported in association with the field action.Additionally, the investigation of this incident previously conducted by engage surgical has been re-opened for review.A supplemental report will be provided upon completion of further investigation activities by smith & nephew, inc.
 
Event or Problem Description
It was reported that, six months after unicompartmental knee arthroplasty was performed on an unknown date, the patient experienced pain.This adverse event was treated with a revision surgery to tka on (b)(6)2021.An abnormally high posterior slope of 19 degrees (the standard is 5 degrees) was produced upon tibial resection, which is the probable root cause the reported pain.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: results of investigation: it was reported that a surgical resident was allowed to perform the tibial resection step on the patient.The surgical resident had not been trained on the proper engage surgical technique or use of the instruments and was not certified by engage to use the system.The attending surgeon reported that a very ¿aggressive¿ tibial cut was made by the resident at the time of surgery.Postop x-ray revealed a super-physiologic posterior slope of 19 deg was measured when used properly the instruments produce a 5 deg nominal posterior slope, which has been shown to be clinically successful in other partial knee replacement systems.An abnormally high posterior slope would lead to elevated tibial strains being introduced into the bone as well as additional stresses being applied to the acl.The x-rays revealed no failure or malfunction of the device.Devices appear well-approximated to the bone with normal in-growth evident at 7-mo on ct evaluation.There was no injury to the patient related to the device, as the revision was due to use by an un-trained user.The patient was successfully revised to a tka at 8-months post-op.The surgeon and rep reported that the implants were well-fixed to the bone prior to removal.Investigation update: as correction action a voluntary market removal will be performed for the engage cementless partial knee system due recent complaint data that indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries.At this time, we have no reason to suspect that the products failed to meet any specifications at the time of manufacture.Based on this investigation, the need for corrective action may be indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POROUS TIBIAL TRAY SIZE 2-RIGHT MEDIAL
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, UNICOMPARTMENTAL/UNICONDYLAR,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17014940
Report Number1020279-2023-01159
Device Sequence Number7031748
Product Code NJD
UDI-Device Identifier00885556873144
UDI-Public885556873144
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K190439
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/09/2025
Device Catalogue Number1-10012-250
Device Lot Number100048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2023
Supplement Date Received by Manufacturer05/26/2023
Initial Report FDA Received Date05/26/2023
Supplement Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 92286
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-