| Model Number |
6ML |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
05/02/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The build lhr for cdm-635l 1582422 aeo861 was examined including the final inspection records and the in-process measurements.There was no ncmr associated with this lot.The device met the m6-c product specification including the axial stiffness and flexural resistance specifications.A review of the risk management files resulted in no new risk.Rmf 0003 rev 36 line 12.58.- infection, infected abscess or infected cyst (not involving resorption or osteolysis), leading to surgical/major intervention with explantation.Rmf 0003 rev 36 line 12.75 - device explanted it should be noted that the company's implants are provided sterile and are validated to an sal of 10-6.There have been no instances of unsterile product release to the field.
|
| |
|
Event or Problem Description
|
|
Information provided states that patient had an m6-c artificial cervical disc implanted at c5/6 in (b)(6) 2022.A revision surgery was performed to explant the device due to patient incurring an infection.
|
| |
|
Search Alerts/Recalls
|