| Model Number |
470405-06 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted ileal conduit surgical procedure, the customer noted the force bipolar instruments jaws were not closing properly.No fragment fell into the patient.The procedure was completed with no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed/replicated the reported complaint.The instrument was found to have a severely bent grip, causing side to side misalignment of the grips.There was a 0.0305¿-0.0350" offset at the tips.This failure is most commonly caused by mishandling/misuse, such as excess force applied to the instrument jaws.Visual inspection was performed and the instrument's grip tips, specifically the gray material, was found to have abrasions measuring roughly 0.0305"-0.0715".The instrument was found to have various scratch marks with light material removed on the main tube.The scratch marks were 0.1300 - 0.2520 in length and were not aligned with the tube axis.This failure is most commonly caused by mishandling/misuse.A broken piece of the gray material measuring roughly 0.0305"-0.0715" was not returned.
|
| |
|
Search Alerts/Recalls
|