| Model Number |
470400-09 |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis investigation confirmed the reported failure.Failure analysis (fa) found a broken conductor wire at the yaw pulley.The conductor wire was broken at the grip-crimp location.The instrument failed the electrical continuity test.No signs of thermal damage were observed.Moreover, the instrument was found to have a broken bipolar yaw pulley.Broken piece is missing because of breakage.Size of missing piece is approximately.059¿ x.092¿.This failure is most commonly caused by mishandling/misuse, such as excess force applied to the distal end of the instrument.
|
| |
|
Event or Problem Description
|
|
It was reported during a da vinci-assisted procedure the long bipolar grasper instrument was not working when the surgical staff tried to use bipolar energy.When the instrument was pulled out a scrub technician found the wire on the instrument snapped.The site completed the procedure as planned with no reported patient injury.
|
| |
|
Search Alerts/Recalls
|