• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KL
Device Problem Priming Problem (4040)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2023
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the customer reported that there was an issue during the priming process.Troubleshooting was performed and found that the customer was not able to exit the load reservoir process/preparing to prime loop.The customer did not receive "complete" status with "next" highlighted on the screen.No harm requiring medical intervention was reported.The customer will discontinue the use of the device and revert to the backup plan as per healthcare professional instruction.The pump will be returned for analysis.
 
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
The pump passed the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, and self test.Test p-cap and reservoir locks properly into the reservoir compartment during testing.No prime/fill anomaly was noted during testing.Pump was cut open to perform visual inspection and found no evidence of physical or moisture damage on the electronic assembly, motor, and force sensor.Motor was tested outside the pump case on the ngp system test board and passed.The following were also noted during visual inspection: scratched case, stained keypad overlay, and pillowing keypad overlay.Prime/fill anomaly was not confirmed during testing.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
justin ellis
ceiba norte ind. park #50 road
juncos 00777--386
7635265677
MDR Report Key17016517
MDR Text Key316094132
Report Number2032227-2023-210660
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000316655
UDI-Public(01)000000763000316655(17)260115
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberNG3323075H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received07/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/16/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age81 YR
Patient SexFemale
Patient Weight61 KG
-
-