| Model Number |
PSO-VT |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Error in pressure measurement at any time for no apparent reason.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis: after opening the dongle shell, there are no cut wires on the circuit board.A pull on the micro-wire inserted in the pa tube shows that it is broken in the tube.Conclusions: the return catheter does not comply with our specifications.Abnormal elongation of the pa tube into which the micro-wire is inserted caused the micro-wire to break internally, resulting in an irreversible electrical discontinuity (e001).Unknown root cause.
|
| |
|
Event or Problem Description
|
|
Error in pressure measurement at any time for no apparent reason.
|
| |
|