MAUDE Adverse Event Report: SOPHYSA PRESSIO® ICP & ICT MONITORING KIT, PARENCHYMAL TUNNELING; PRESSIO® INTRACRANIAL PRESSURE MONITORING KIT, VENTRICULAR TUNNELING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOPHYSA PRESSIO® ICP & ICT MONITORING KIT, PARENCHYMAL TUNNELING; PRESSIO® INTRACRANIAL PRESSURE MONITORING KIT, VENTRICULAR TUNNELING Back to Search Results
Model Number PSO-VT
Medical Device Problem Codes Obstruction of Flow (2423); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2023
Type of Reportable Event Malfunction
Event or Problem Description
Error in pressure measurement at any time for no apparent reason.
 
Additional Manufacturer Narrative
Analysis: after opening the dongle shell, there are no cut wires on the circuit board.A pull on the micro-wire inserted in the pa tube shows that it is broken in the tube.Conclusions: the return catheter does not comply with our specifications.Abnormal elongation of the pa tube into which the micro-wire is inserted caused the micro-wire to break internally, resulting in an irreversible electrical discontinuity (e001).Unknown root cause.
 
Event or Problem Description
Error in pressure measurement at any time for no apparent reason.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSIO® ICP & ICT MONITORING KIT, PARENCHYMAL TUNNELING
Common Device Name
PRESSIO® INTRACRANIAL PRESSURE MONITORING KIT, VENTRICULAR TUNNELING
Manufacturer (Section D)
SOPHYSA
5 rue guy moquet
orsay, 91400
EN  91400
Manufacturer (Section G)
SOPHYSA
rue sophie germain
besançon, 25000
FR   25000
Manufacturer Contact
marion prud'homme
5 rue guy moquet
orsay, 91400
FR   91400
MDR Report Key17017175
Report Number3001587388-2023-23101
Device Sequence Number7126454
Product Code GWM
UDI-Device Identifier03760124131482
UDI-Public(01)03760124131482(11)220922(17)270601
Combination Product (Y/N)N
Initial Reporter CountryBE
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial,Followup
Report Date (Section B) 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberPSO-VT
Device Catalogue NumberPSO-VT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2023
Supplement Date Received by Manufacturer03/01/2023
Initial Report FDA Received Date05/29/2023
Supplement Report FDA Received Date01/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age47 YR
Patient SexFemale
-
-