| Model Number |
110-120 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/29/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported right side "rupture." the device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued h6 health effect impact code: f2203- imaging required.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "rupture." the device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "rupture." the device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed, a broken in the posterior side with non-penetrating nicks assessed as to surgical impact (shell thickness was within specification).Also observed in the device a opening striated assessed as to surgical damage.Additional observations: ¿ yellow particles observed in the device.¿ crease flat observed in the device.No further actions are required as the device was implanted.
|
| |
|