| Model Number |
MI1250 SYNCHRONY 2 |
| Medical Device Problem Codes |
Circuit Failure (1089); No Device Output (1435)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023 the user was seen in the clinic and there was no communication with the processor.The link check would only flash red.With troubleshooting and providing replacement devices (external), the problem still could not be remedied.The user was seen again on (b)(6) 2023 and the problem still persisted.Re-implantation has been scheduled.
|
| |
|
Event or Problem Description
|
|
On (b)(6), 2023 the user was seen in the clinic and there was no communication with the processor.The link check would only flash red.The user was seen again on (b)(6), 2023 and the problem still persisted.The user had no more benefit with the device.The user was re-implanted on (b)(6), 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusions: device investigations revealed the substrate of the device circuitry to be broken.The investigation results appear to match the problems mentioned in the recipient report.The exact reason for the crack in the substrate cannot be determined.However, a review of the dhr has been performed and it was confirmed that the device was released according to specifications.Furthermore, before the occurrence of the above issue, two affected channels had been observed due to unknown reasons.This is a final report.
|
| |
|