Model Number EIS-HCF25
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was received and evaluated.Device evaluation confirmed the reported issue.Inspection found damaged tip, noted to be sharp.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event or Problem Description
Damaged tip was reported on the device.The issue found during reprocessing.There was no patient harm, no user injury reported.Device return evaluation found broken tip , sharp.This report is being submitted due to the finding of broken tip , sharp identified during device evaluation.
 
Additional Manufacturer Narrative
Updated: g2, h4, h6, h10 this report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The manufacturing date for this product lot is april 2016.The final manufacturing date was not provided in the dhr, therefore, h4 has been update to apr 1, 2016.Based on the results of the investigation, a definitive root cause of the reported damaged/broken inner sheath tip issue could not be determined, however, the issue was likely the result of user mishandling, such as such as device impact/dropped during device reprocessing.The event may be detected/prevented by following the instructions for use which state: "study this manual and other labeling thoroughly for safe handling, storage and usage, including instructions for all generators and accessories.Failure to properly follow the instructions, warnings, and cautions may lead to serious surgical consequences or injury to the patient.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects.Do not use an instrument that fails to meet the criteria stated in the labeling or that has been damaged.Damage may result in the loss of the entire ceramic tip or fragments of the ceramic tip.If there is evidence of charring, burn spots, chips or cracks in the ceramic tip or surrounding area, do not use." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CF RESECTOSCOPE INNER SHEATH, 25FR
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
GYRUS ACMI, INC.
800 west park drive
westborough MA 01581
Manufacturer (Section G)
GYRUS ACMI, INC.
800 west park drive
westborough MA 01581
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17021076
Report Number3003790304-2023-00203
Device Sequence Number15344920
Product Code HIH
UDI-Device Identifier00821925001893
UDI-Public00821925001893
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K890328
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEIS-HCF25
Device Lot NumberRXDW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2023
Supplement Date Received by Manufacturer06/20/2023
Initial Report FDA Received Date05/30/2023
Supplement Report FDA Received Date07/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-