| Catalog Number |
109990 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during treatment with continuous renal replacement therapy (crrt) using a prismaflex control unit, a prismaflex m150 set, an alarm condition related to 'air in blood' was generated and ¿clots throughout the entire deaeration chamber and clots at the bottom¿ that were preventing the air from being removed from the return line were noted.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|