Model Number 3501
Device Problems Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581); Low impedance (2285)
Patient Problem Electric Shock (2554)
Event Date 07/19/2021
Event Type  Injury  
Event Description
It was reported that the patient had an inappropriate shock due to oversensing.The shock occurred in the secondary sensing vector.The device has now been reprogrammed to the alternate sensing vector.The shock impedance was less than 30 ohms.The clinic stated the patient is a very frail gentleman.The patient will be monitored via latitude.There were no additional adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17025968
MDR Text Key316193426
Report Number2124215-2023-27325
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526599200
UDI-Public00802526599200
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/13/2021
Device Model Number3501
Device Catalogue Number3501
Device Lot Number154578
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient SexMale
-
-