| Model Number |
212186 |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Worn saw attachment- replacement needed unable to use - rust case type / application: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.H3 other text : device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event.An event regarding corrosion involving a mako saw attachment was reported.The event was confirmed.Method & results.-product evaluation and results: visual inspection confirmed the reported event.Red colored residue present at shaft connection point to locking assembly -clinician review: no medical records were received for review with a clinical consultant.-product history review: a product history review is not required, as no manufacturing specific issue has been alleged.At stryker joint replacement all devices/product are manufactured through validated equipment and inspected by trained representatives through stryker¿s quality management system.-complaint history review: a complaint history review and trend detection is not required as this is a pm.Monitoring will be continued under the current quarterly trend review.Conclusions: no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|
Event or Problem Description
|
|
Worn saw attachment- replacement needed unable to use - rust.Case type / application: tka.
|
| |
|