• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE; PKL, LIGATOR, HEMORRHOIDAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE; PKL, LIGATOR, HEMORRHOIDAL Back to Search Results
Catalog Number INSC-P-7-230-S
Device Problems Material Separation (1562); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
During an endoscopy, the physician used two cook instinct plus endoscopic clipping devices.It was reported that after the clips were put down the scope, the nurse went to open the clips but they detached from the catheter and stayed attached to the drive wires.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Pma/510(k): k212323.Investigation evaluation: a product evaluation was not performed in response to this report because the products said to be involved were not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: a corrective action has been initiated to reduce occurrences of the clip housing detaching from the cath attach for instinct plus endoscopic clipping devices.The product said to be involved is included in the scope of the corrective actions.The instructions for use states: "uncoil device.Verify smooth handle operation and clip action.Open clip by gently moving handle spool distally (away from handle thumb ring).Once clip is fully open, do not continue advancing handle spool as clip may prematurely detach from catheter.Close clip by moving handle spool proximally until clip is fully closed.Precaution: do not continue to pull handle spool beyond tactile resistance as this may prematurely deploy clip." the instructions for use states: "with clip closed and without holding handle spool, advance device in small increments into accessory channel of gastroscope, duodenoscope, or colonoscope.Note: if difficult to advance device, relax elevator and/or straighten endoscope.¿ the ifu also includes the following precaution: ¿holding handle spool during clip advancement may prematurely deploy clip.¿ failure to follow the instructions above can result in damage to the device which may lead to premature clip deployment.Prior to distribution, all instinct plus endoscopic clipping devices are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Type of Device
PKL, LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key17030540
MDR Text Key316702746
Report Number1037905-2023-00260
Device Sequence Number1
Product Code PKL
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberINSC-P-7-230-S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DUODENOSCOPE, UNKNOWN MAKE AND MODEL
-
-