| Model Number |
68LP-400 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported a right side deflation and "squishy, smaller".Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade i.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Patient reported a right side deflation and "squishy, smaller".Healthcare professional reported right side capsular contracture, baker grade i.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed an opening on posterior assessed as fold flaw opening.¿ capsular contracture: unable to observe since it is not related to the device.¿ visibility/palpability: unable to observe since it is not related to the device.Additional observations: ¿ crease fold and wear abrasion observed on the device.¿ yellow particles observed inside the device.No further actions are required as the device was implanted.
|
| |
|