• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMBLEM MRI S-ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMBLEM MRI S-ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number A219
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Patient Problem Electric Shock (2554)
Event Date 05/24/2023
Event Type  malfunction  
Event Description
It was reported that the patient had an inappropriate shock due to oversensing of myopotential noise via a change in the intrinsic morphology.Technical services discussed optimizing the sensing.There were no additional adverse patient effects.The system remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM MRI S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17031798
MDR Text Key316234518
Report Number2124215-2023-27511
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526584404
UDI-Public00802526584404
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/22/2022
Device Model NumberA219
Device Catalogue NumberA219
Device Lot Number259666
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2023
Initial Date FDA Received05/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-