Model Number MC2-5090S
Medical Device Problem Codes Break (1069); Corroded (1131); Fracture (1260); Mechanical Jam (2983)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Event or Problem Description
An investigation report from an implant retrieval center was received and reported that during examination, the rod would not distract and force testing was unable to be performed as the electronic remote controller (erc) was not functional.Additionally, it was noted during surface damage grading that there was corrosion.It was reported that the rod was removed due to a locking pin fracture.There was no adverse patient or user impact associated with this report.No additional information is available.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.The evaluation was performed by the london implant retrieval center (lirc).The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.Report 1 of 2.
 
Event or Problem Description
No additional information has been provided.
 
Additional Manufacturer Narrative
Corrected data: h6 (health effect - impact code).The device was not returned to nuvasive for evaluation, however, the provided patient x-rays were reviewed and confirmed the locking pin failure.Without the return of the device, the root cause is unable to be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC GENERATION 2B SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr. ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key17032141
Report Number3006179046-2023-00239
Device Sequence Number7012406
Product Code PGN
UDI-Device Identifier00887517834041
UDI-Public887517834041
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5090S
Device Lot Number7092214AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2023
Supplement Date Received by Manufacturer08/24/2023
Initial Report FDA Received Date05/31/2023
Supplement Report FDA Received Date08/30/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age9 YR
Patient SexMale
-
-