| Model Number |
MC2-5070S |
| Medical Device Problem Codes |
Break (1069); Corroded (1131); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The evaluation was performed by the london implant retrieval center (lirc).The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
An investigation report from an implant retrieval center was received and reported that the rod passed force testing.Additionally, it was noted during surface damage grading that there was galling and corrosion.It was reported that the rod was removed due to it being broken; there was no adverse patient or user impact associated with this report.No additional information is available.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to nuvasive for evaluation, however, the provided patient x-rays were reviewed and confirmed a solid section break.Without the return of the device, the root cause is unable to be determined.
|
| |
|