Model Number MX60ET
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 04/26/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported one day after implantation of an intraocular lens into the left eye (os) the lens rotated out of position.Approximately one-month post-op, the iol was repositioned to its intended position.In the surgeon''s opinion, the cause of the iol rotation is unknown.Patient outcome is good.
 
Additional Manufacturer Narrative
The lens remains implanted and is not available for investigation.A review of the device history record (dhr) did not identify any anomalies or nonconformities that could be related to this event.The lot history, trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause of this event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVISTA TORIC INTRAOCULAR LENS
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
21 north park place blvd.
clearwater FL 33759
Manufacturer Contact
shayan habibi
21 n park pl blvd.
clearwater, FL 33759
7277246600
MDR Report Key17032574
Report Number0001313525-2023-70064
Device Sequence Number6516124
Product Code MJP
UDI-Device Identifier10757770555649
UDI-Public(01)10757770555649(17)250331
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P910056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model NumberMX60ET
Device Catalogue NumberMXUET200+180
Device Lot Number3247731
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BAUSCH AND LOMB INJECTOR SYSTEM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-