Model Number PM3210
Medical Device Problem Codes Pacemaker Found in Back-Up Mode (1440); Therapy Delivered to Incorrect Body Area (1508); Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented to in clinic follow up with diaphragm stimulation.The pacemaker was found to be in backup mode after an attempted cybersecurity upgrade.The device was successfully restored back to normal settings.The patient was stable.
 
Additional Manufacturer Narrative
Correction: h6 coding for backup reset.
 
Additional Manufacturer Narrative
Correction: h6 coding should have been pacemaker found in back up mode not unexpected back up mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM RF CRT-P
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17033130
Report Number2017865-2023-22068
Device Sequence Number7012028
Product Code NKE
UDI-Device Identifier05414734503785
UDI-Public05414734503785
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2015
Device Model NumberPM3210
Device Catalogue NumberPM3210
Device Lot Number4493400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2023
Supplement Date Received by Manufacturer09/21/2023
01/10/2024
Initial Report FDA Received Date05/31/2023
Supplement Report FDA Received Date09/21/2023
01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/27/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDTRONIC LEAD; QUICKFLEX
Patient Age75 YR
Patient SexFemale
-
-