• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLT, PLS, AUTO RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLT, PLS, AUTO RBC SET Back to Search Results
Model Number 82656
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: the analysis of the run data file did not identify a conclusive root cause for the higher-than-expected wbc content reported in the platelet product for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.The signals indicate that the device had difficulty maintaining steady state operation during the procedure, so it cannot be ruled out that the needle line access may have shifted and could potentially have impacted the collection.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product may have been donor-related.
 
Event or Problem Description
The customer has reported an unflagged wbc contamination issue.The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit #: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL PLT, PLS, AUTO RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key17033780
Report Number1722028-2023-00192
Device Sequence Number20436234
Product Code GKT
UDI-Device Identifier05020583826564
UDI-Public05020583826564
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2024
Device Model Number82656
Device Catalogue Number4826562
Device Lot Number2211122142
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2023
Initial Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-