| Catalog Number |
0616002 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hernia (2240)
|
| Date of Event |
10/09/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This legal event is being reported as serious injury due to the reported recurrence with surgical intervention.The internal investigation is pending and will be reported in a follow up report along with the investigation conclusion.
|
| |
|
Event or Problem Description
|
|
It was reported through a legal event that a 45 year old female had hernia repair surgery on or about (b)(6) 2016.During the hernia repair surgery, the surgeon implanted a strattice mesh; lot # sp100326-060; ref # 0616002) mesh.After surgery, the patient returned to the hospital on or about (b)(6) 2016, for a revision surgery.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional and/or corrected data: b5, d4, g6, h2, h6 internal investigation into strattice lot sp100326 included a review of the reported information, review of the device history records, and a review of the complaint history records.The investigation resulted in no remarkable findings, including no other complaints reported against the lot and no deviations or related non-conformances revealed during processing.The lot was terminally sterilized within the process parameters and met all qc release criteria.Based on our internal investigation with no remarkable findings, and without relevant patient factors, a relationship between the strattice and this event could not be determined.Due to the legal process, if additional information is made available during legal proceedings, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
This is follow up#1 to report the results from the internal investigation and the conclusion.As reported in the initial: it was reported through a legal event that a 45 year old female had hernia repair surgery on or about (b)(6) 2016.During the hernia repair surgery, the surgeon implanted a strattice mesh; lot # sp100326-060; ref # 0616002) mesh.After surgery, the patient returned to the hospital on or about (b)(6) 2016, for a revision surgery.No other information was provided.
|
| |
|
Search Alerts/Recalls
|