• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. SET SCREW DRIVER; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. SET SCREW DRIVER; SCREWDRIVER Back to Search Results
Catalog Number 230H4001
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that two final drivers would not retain the mating set screws intra-operatively.There were no impacts to the patient.However, device evaluation by the manufacturer found that the tips had fractured off both drivers.This is report one of two for this event.
 
Additional Manufacturer Narrative
Device evaluation -the returned devices match the information in the complaint file and were examined.Visual inspection revealed that the spring had a kink in it and the head of the device was also tilted.The set screw drivers had fractured tips.See images attached in the sub-form.Potential root cause -a definitive root cause cannot be determined with the information provided.These events could possibly be attributed to wear through use over time or multiple sterilization cycles.These events could also be attributed to off-axis forces applied during use.Dhr review -the dhrs were reviewed.There are no indications of manufacturing issues which would have contributed to these events and the devices were likely conforming when they left zimvie¿s control.Device usage -this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.Reference report 3012447612-2023-00175.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET SCREW DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key17035107
Report Number3012447612-2023-00174
Device Sequence Number12391178
Product Code HXX
UDI-Device Identifier00887868227431
UDI-Public(01)00887868227431(10)313114
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number230H4001
Device Lot Number313114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2023
Initial Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-