STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Model Number |
6021-0335 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646)
|
| Date of Event |
05/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding loosening involving an accolade stem was reported.The event was not confirmed.Method & results: product evaluation and results: the reported device was not returned however photographs were provided for review.The photographs show a recently explanted stem with blood and tissue visible on the trunnion and stem neck.The device appears otherwise unremarkable.Material analysis, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to stem loosening.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text : not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left hip was revised due to pain and aseptic loosening of the femoral stem.The stem, ceramic head, and competitor liner were revised to a restoration modular stem construct with a metal head and another competitor liner.
|
| |
|
Search Alerts/Recalls
|
|
|