• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number PXT281212
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 01/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2010, this patient underwent an endovascular treatment for an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis.A trunk - ipsilateral leg endoprosthesis (pxt281212) was deployed, and its ipsilateral leg was extended by a contralateral leg endoprosthesis (pxc181000).Another contralateral leg endoprosthesis (pxc201400) was deployed as a contralateral limb.As the pxc201400 landed on a tightly curved left common iliac artery, it was reinforced by a cordis palmaz® stent.The final angiography confirmed a type endoleak, which was decided to be monitored and left untreated.The patient tolerated the procedure.(ref.(b)(4)) on (b)(6) 2015, reintervention was performed to treat a proximal type endoleak.While securing the left renal flow by using a boston scientific express¿ vascular sd as a chimney graft, an aortic extender endoprosthesis was added to the proximal end of the previously implanted pxt281212.The final angiography confirmed no endoleaks.The patient tolerated the procedure.(ref.(b)(4)) on an unknown date around 2021, a follow-up examination confirmed an unknown amount of aneurysm enlargement.As treatment, the patient underwent an open vascular graft replacement during which the previously implanted pxt281212, pxc181000 and pxc201400 were partially explanted.On an unknown date, a follow-up examination confirmed a distal type : endoleak from the left contralateral limb where the pxc201400 had been implanted.On (b)(6) 2023, reintervention was performed to treat the observed distal type : endoleak.After coil embolizing the left internal iliac artery with abbott amplatzer¿ vascular plug 4, two contralateral leg endoprosthesis (plc121400j+plc121000j) were added to the distal end of the left contralateral limb.The endoleak was resolved.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
H3: code "other" was selected as the medical device remains implanted.Return not possible.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Emdr section h6 codes b, c, d were updated to reflect results of investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY MOUNTAIN B/P
3750 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
samir kulovic
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key17037059
Report Number2017233-2023-03971
Device Sequence Number17895657
Product Code MIH
UDI-Device Identifier00733132610259
UDI-Public00733132610259
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2012
Device Model NumberPXT281212
Device Catalogue NumberPXT281212
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2023
Supplement Date Received by Manufacturer05/08/2023
Initial Report FDA Received Date06/01/2023
Supplement Report FDA Received Date09/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/26/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-