It was reported that healing color does not enter.After the tsv implant was placed, the doctor tried to apply a healing collar during the secondary operation, but it was reported that it did not work.The doctor tried hc343 and tried hc345.However, both did not enter.Both were new.The doctor entered with another hc, so it was completed.No health hazard.The doctor has doubts and wants a survey response of the actual product.
|
|
Zimvie received one (1) hc345, (heal collar 3.5x4.5, 5mm) for evaluation.Visual evaluation was performed, a damage threads identified resulting in failure to seat.Dhr review was completed for the subject lot number 2020091295.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2020091295 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : fit : does not seat.The customer did not submit images for the reported event.Based on the investigation and risk management file review as per rm-00057-haz, the most likely root cause determined from the investigation was missing or confusing instructions for use and torque applied during placement/ seating exceeds recommended value.Therefore, based on the available information, a device malfunction did occur.The abutment had damage threads identified resulting in failure to seat.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: g6: checked "follow-up." h3: changed "no" to "yes.".
|