Model Number ACU0T0
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A health care professional reported that, during the procedure, lens became lodged in the injector / introducer and was unable to be used and implanted.The lens was discarded and new lens and introducer obtained and used.Additional information has been requested.
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.It is unknown if a qualified viscoelastic was used.The instruction for use (ifu) instructs: during device preparation and implantation of the company iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Lens may become stuck in the device: ¿ if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to become ¿stuck¿ in the device ¿ due to the use of a non-qualified viscoelastic.Material properties of non-qualified ophthalmic viscosurgical devices (ovds), may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.¿ if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17038286
Report Number1119421-2023-01016
Device Sequence Number6908986
Product Code HQL
UDI-Device Identifier00380652395250
UDI-Public00380652395250
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/17/2025
Device Model NumberACU0T0
Device Lot Number15470729
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/05/2023
Supplement Date Received by Manufacturer07/18/2023
Initial Report FDA Received Date06/01/2023
Supplement Report FDA Received Date08/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/19/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-