(b)(4).D10: 423859 5mm reverse curette 147260.G2: foreign: netherlands.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2023 - 01253.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Updated: d4, d9, g3, g6, h2, h3, h4, h6 proposed component code mechanical (g04) - drill an ultra drive 7mm disk drill, part # 423871 from lot 600710, was returned and evaluated against the complaint.The packaging lot was confirmed with the complaint.Nicks and scratching are present on the proximal end of the drill near the connection feature.Scratching and scuffing are present on the shaft.The disk drill tip is fractured and was not returned with the remainder of the device.A review of the device history records identified no deviations or anomalies during manufacturing.A review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.A definitive root cause cannot be determined.Event is confirmed via returned product.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|