DEPUY IRELAND - 3015516266 ATT REV SLV LPS FEM ADP +0; KNEE FEMORAL ADAPTOR Back to Search Results
Model Number 1511-06-000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscular Rigidity (1968); Loss of Range of Motion (2032); Joint Laxity (4526)
Date of Event 05/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had poor motion post total knee and some issues with patellar tracking.Joint line had been established by previous surgery before the surgeon had operated on her and the tibia was unable to be removed so he chose to revise the femoral side to reduce the tension in her joint.Femoral side implants were removed and the bone was prepared and new implants were placed slightly more proximal allowing better motion and less tension on the lateral retinacular structures.No instruments were broken nor patient harm occurred.New pin from new femoral component was utilized.Doi: (b)(6) 2023.Dor: (b)(6) 2023.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received: a.Was instability related to polyethylene wear? --> no b.Was there any reported malposition of components that led to the instability? --> no c.Were any components undersized on the primary surgery that led to the instability? --> no d.Was the femur or tibia excessively resected/joint line raised during the primary surgery that led to the instability? --> joint line was set by previous surgery.E.Please clarify if the patient's patella had already resurfaced prior to this revision, and if so, was the patella product a depuy patella implant? if it is, it will need to be addressed for joint instability for the patella tracking issues.--> it had an unidentifiable implant in place and remained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATT REV SLV LPS FEM ADP +0
Common Device Name
KNEE FEMORAL ADAPTOR
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17038469
Report Number1818910-2023-11299
Device Sequence Number15067710
Product Code MBH
UDI-Device Identifier10603295503408
UDI-Public10603295503408
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K202248
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1511-06-000
Device Catalogue Number151106000
Device Lot NumberJC8949
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2023
Supplement Date Received by Manufacturer06/01/2023
06/12/2023
Initial Report FDA Received Date06/01/2023
Supplement Report FDA Received Date06/06/2023
06/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATUN FEM SLV M/L 30MM FULL POR; ATUN PRESSFIT STR STEM16X110MM; CONTROL CABLE W/CABLE SLEEVE; HP STRL THREADED PINS; LPS DISTAL FEM COMP XXSM RT; LPS UNIV TIB HIN INS XXSM 12MM; UNK ATTUNE KNEE PATELLA
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR