• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED UTICA ELEC,BLDE,LNG,EXT,INSULATD,40; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED UTICA ELEC,BLDE,LNG,EXT,INSULATD,40; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 138110
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
During incoming inspection, the distributor rejected this device, 138110, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
Reported event of an insufficient heatseal is confirmed.Received one 138110 in original packaging.Lot number was verified.Performed a visual inspection of the device there were no obvious signs of a breach in the seal however, the device was encroaching into the seal which broke the heat seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised the following: sterility guaranteed unless package has been opened, broken, or damaged.Inspect and test each device before each use.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEC,BLDE,LNG,EXT,INSULATD,40
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key17038560
Report Number1320894-2023-00104
Device Sequence Number6908563
Product Code GEI
UDI-Device Identifier30653405003452
UDI-Public(01)30653405003452(17)270103(10)20220141
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943542
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number138110
Device Catalogue Number138110
Device Lot Number202201041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2023
Initial Report FDA Received Date06/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-