| Model Number |
PNDL, 5BV TVH TP 32G 4MM100CT 30/CS |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Consumer reported complaint for the trueplus pen needles.Consumer reported complaint she is unable to get the insulin out of the 32g pen needles.The customer has been using this product for several weeks.The package had not been open or damaged when received by the customer.The customer is using compatible product and the pen needle is properly aligned.At the time of the call the customer felt well and did not report any symptoms.Customer did not claim to be injured while using the pen needles and no medical intervention related to the use of the product was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Pen needles were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on 17-may-2023 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 20-jul-2023: h6: updated fda¿s type, findings and conclusions codes.H10: pen needles were not returned for evaluation.Complaint was forwarded to supplier quality and internal evaluation was performed by the manufacturer using pen needles from the same lot.No abnormalities observed with retention samples.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
|
| |
|