MENTOR TEXAS MENTOR MEMORYSHAPE¿ MM+ 295CC; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3341155 |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cyst(s) (1800)
|
| Date of Event |
05/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: cyst.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent primary augmentation surgery with mentor glow study gel devices on (b)(6) 2020.Adverse event # 1.Breast cyst, (left side, implant serial number: (b)(6).Doi: (b)(6) 2020).On (b)(6) 2020, she presented with breast cyst, which required cyst aspiration on (b)(6) 2020.The principal investigator assessed this event as mild in severity, non-serious, not related to the device.Should additional information become available, this case will be re-assessed and notes will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Per additional information received on december 18, 2025, the patient scheduled for removal surgery on (b)(6) 2026, patients date of birth is (b)(6) 1979, procedure was breast augmentation primary, patient ethnicity is white, and non hispanic, mri confirmed interacapsular rupture, and was performed on (b)(6) 2025.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Adverse event # 2.False positive rupture per imaging (left side, implant serial number: (b)(6).Doi: (b)(6) 2020, doe: (b)(6) 2026).On (b)(6) 2020, she presented with false positive rupture per imaging, which required the bilateral replacements per following: r) smpb290 sn: (b)(6), l) smpb-290 sn: (b)(6), on (b)(6) 2026.The principal investigator assessed this event as moderate in severity, serious, definitely related to the device.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on march 25, 2026: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the returned device.A breast cyst is a fluid-filled sac within the breast.One breast can have one or more breast cysts.They are often described as round or oval lumps with distinct edges.Breast cysts are usually and typically do not require treatment, but they can be drained using fine needle aspiration if the cyst is large or uncomfortable.The contralateral device was also received (lot number 7540883).The patient did not report any issue with this device.Therefore, no further analysis of the contralateral device is required.As part of mentor¿s quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Manufacturer¿s reference number: (b)(4).
|
| |
|