| Model Number |
0220220300 |
| Medical Device Problem Codes |
Device Sensing Problem (2917); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib, mike, biomed, the light went dim and would not brighten, temp the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.Although the complaint could not be confirmed, the probable root cause could be a bad reed switch in the safe light cable used at the time of the case.Additionally, one of the led clips on the main board are open which can cause issues with the lightsource's ability to maintain the proper color balance.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|