• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number A2030-120
Medical Device Problem Codes Leak/Splash (1354); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a lesion in an unspecified artery.The 3.0x120mm armada 14 balloon dilatation catheter (bdc) had issues when trying to screw the inflation device on the catheter.Upon fixing the inflation device manually against the balloon lumen, there was no pressure build up.Another balloon was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
Visual and functional inspections were performed on the returned device.The reported loose or intermittent connection or leak of the hub could not be confirmed; however, a pinhole rupture on the balloon was noted.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaints or note balloon rupture.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17040036
Report Number2024168-2023-05733
Device Sequence Number6459560
Product Code DQY
UDI-Device Identifier08717648156106
UDI-Public08717648156106
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model NumberA2030-120
Device Catalogue NumberA2030-120
Device Lot Number2102541
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/10/2023
Supplement Date Received by Manufacturer07/20/2023
Initial Report FDA Received Date06/01/2023
Supplement Report FDA Received Date07/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-