WALDEMAR LINK GMBH & CO. KG LINKSYMPHOKNEE; ARTICULATING SURFACE, SYMMETRIC, SZ. 7-8, UHMWPE PS, FIXED, H= 12 MM
|
Back to Search Results |
|
| Model Number |
880-217/12 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The review of the device history record showed no deviations.The product complies with the specifications valid at the time of manufacture.The report is delayed due to original classification of the case as non-reportable.Due to the recent fda inspection we reevaluated the complaint in comparison to similar complaints with different outcomes and therefore determined that it is reportable.We have addressed this observation in capa-23-08.
|
| |
|
Event or Problem Description
|
|
Lb received and complaint sample and information on (b)(6) 2022, from (b)(6), regarding a ps poly insert that "would not engage to tibial tray.Poly was not used, we opened another poly insert that worked fine with tibial base plate".[customer].
|
| |
|
Search Alerts/Recalls
|
|
|