• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN¿ LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 2MM(H); DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN¿ LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 2MM(H); DENTAL IMPLANT Back to Search Results
Catalog Number ILPC442U
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
Doctor reported that patient came with loose inferior all on four implant system at tooth sites 32 - 42 and noticed that the low profile screws were fractured.New low profile screws were placed to complete the procedure.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Zimvie received one (1) ilpc442u, (certain low profile one-piece abutment 4.1mm(d) x 2mm(h)) for evaluation.Visual evaluation was performed, a fracture has been identified at the threads.Dhr review was completed for the subject lot number 2022110355.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2022110355 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : screw.Based on the investigation and risk management file review, the most likely root cause determined from the investigation were missing or confusing instructions for use and clinician incorrectly engages abutment screw into implant.Therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the threads.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN¿ LOW PROFILE ONE-PIECE ABUTMENT 4.1MM(D) X 2MM(H)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key17043387
Report Number0001038806-2023-01052
Device Sequence Number17425725
Product Code NHA
UDI-Device Identifier00844868021063
UDI-Public(01)00844868021063(17)271202(10)2022110355
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K092341
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberILPC442U
Device Lot Number2022110355
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/12/2023
Supplement Date Received by Manufacturer05/12/2023
Initial Report FDA Received Date06/01/2023
Supplement Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age73 YR
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-