| Model Number |
02-01650 |
| Medical Device Problem Codes |
Overheating of Device (1437); Temperature Problem (3022); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was hot to the touch and the battery was swelling/budging with visible cracks.There was no report of death or serious injury, or impact to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Engineering evaluation was completed on 22may2023.Device was inspected for general physical integrity, the device case was cracked.Device started charge at 9:10am at 65.5ma, slow blinking white led, at 79.4f.Device charging was stopped at 9:40am at 449.3ma at 107.4f as this exceeds the limit for device heating.See attachment 1 for results.Device was opened to visually inspect for damage and corrosion that may have caused of contributed to allegation.The printed circuit board showed no signs of damage.The battery was swollen and the covering was wrinkled.Conclusion: engineering evaluation was able to confirm charging overheat/bulging issue.Device exceeded normal operating temperatures.Root cause is likely due to the swelling of the battery.Summary: it was reported that the device was hot to the touch and the battery was swelling/budging with visible cracks.The device returned for investigation and engineering evaluation was able to confirm charging overheat/bulging issue.Device exceeded normal operating temperature.Root cause is most probable to be due to the swelling of the battery.
|
| |
|