As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 06/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.However, the device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the halo device was returned for evaluation.Two splits were confirmed on the dilator tip.One split was 5mm in length, the second split is 4.5mm in length.The sheath tip was undamaged.The result of the investigation is confirmed for the reported tip split issue.The root cause for the reported tip split issue could not be determined based upon the available information received from the field communications and sample evaluation.Labeling review: instruction for use for the halo one thin-walled guiding sheath was reviewed and the following sections are applicable: indication for use: the halo one¿ thin-walled guiding sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices.The halo one¿ thin-walled guiding sheath is not indicated for use in the neurovasculature or the coronary vasculature.Warnings: 2.Cleaning, reprocessing and/or resterilization of the present medical device increases the probability that the device will malfunction due to potential adverse effects on components that are influenced by thermal and/or mechanical changes.4.Visually inspect the packaging to verify that the sterile barrier is intact.Do not use if the sterile barrier is opened or damaged.6.Do not advance the guidewire, sheath/dilator, procedural device, or any component if resistance is met, without first determining the cause and taking remedial action 10.Only advance or retract the sheath with the dilator inserted and only advance or retract the sheath and dilator while placed over a properly sized guidewire.11.Failure to deactivate the procedural device prior to removal through the sheath may cause damage to the sheath and may result in patient injury.Precautions: 1.The halo one¿ thin-walled guiding sheath shall only be used by trained physicians.Access procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment and / or ultrasound.4.The minimum acceptable sheath french size is printed on the package label.Do not attempt to pass devices through a smaller size sheath introducer than indicated on the device label.6.Carefully inspect the sheath prior to use to verify that the sheath has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.8.Insert dilator into the center of the sheath valve.Forced insertion of the dilator which misses the center of the valve may cause damage and result in blood leakage.9.Advance or withdraw the sheath slowly.If resistance is met do not advance or withdraw until the cause of resistance is determined.10.When inserting, manipulating or withdrawing a device through the introducer always maintain the introducer position.11.Remove the dilator from the sheath slowly to avoid incomplete closing of the valve resulting in blood leakage.12.When using procedural devices close to the tip of the sheath care must be taken to ensure the active mechanism portion of the procedural device (e.G., balloon, stent zone, material removal section of atherectomy device) is not within the tip of the sheath.The radiopaque marker is located within 5 mm of the end of the tip but does not mark the true distal tip of the sheath.13.Before removing or inserting the interventional/diagnostic device through the sheath, aspirate blood from the 3-way stopcock to remove any fibrin deposition which may have accumulated in or on the tip of the sheath.14.Ensure to deactivate the procedural device prior to removal through the sheath.15.Take care when back loading the tip of the dilator over the guidewire to avoid damage to the dilator.18.Do not place sutures on the sheath tubing since this may restrict access/flow through the sheath.When puncturing, suturing or incising near the sheath be careful not to damage the sheath.Proper functioning of the sheath depends on its integrity.Care should be used when handling the sheath.20.If resistance is felt during post-procedure withdrawal of the procedural device, it is recommended to remove the procedural device, guidewire, and sheath as a single unit.22.Proper functioning of the halo one¿ thin-walled sheath depends on its integrity.Care should be used when handling the sheath.Damage may result from kinking, stretching, or forceful wiping of the halo one¿ thin-walled guiding sheath.Do not continue to use the sheath if the shaft has been bent or kinked.Directions for use: 1.Verify the french size is suitable for the procedure and can accommodate the required procedural devices as labeled.Remove sheath and dilator from package.4.Insert the provided vessel dilator through the hemostatic valve and click the dilator hub into place in the valve housing 6.Identify the insertion site, including but not limited to radial, femoral, popliteal, tibial, or pedal access sites and prepare the site using proper aseptic technique and local anesthesia as required.7.At the operator¿s discretion, make a small skin incision at the puncture site with a surgical scalpel.Recommended for scar tissue at the access site.8.Backload the distal tip of the halo one¿thin-walled guiding sheath dilator over the prepositioned guidewire and advance the tip to the introduction site.9.Advance the dilator and the sheath through the skin and into the vessel.Grasp the sheath and dilator assembly close to the skin while advancing to avoid buckling, employ a rotating motion while advancing as required.(note: if using a hydrophilic guidewire, ensure that it is kept hydrated with sterile heparinized saline, at all times.) 10.Carefully advance the dilator and the sheath over the wire to the site required within vessel.The radiopaque marker identifies the sheath tip location under fluoroscopy.11.Disconnect the dilator hub from the valve by bending to the side to unsnap from the valve cap.Remove the dilator slowly while holding the sheath in place and ensuring the guidewire remains in place as required.12.Load the procedural device over the pre-positioned guidewire.13.Advance the procedural device carefully into the centre of the valve diaphragm and through the sheath to the treatment site while maintaining the sheath position.14.Position the procedural device relative to the lesion to be treated ensuring the active mechanism portion of the procedural device is not within the sheath.16.After the procedure is completed insert the dilator over the wire into the sheath 17.Slowly withdraw the sheath and dilator as a unit and then remove the guidewire.H10: d4 (expiry date: 06/2023), g3.H11: h6 (method, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|