| Model Number |
101/815/070 |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the length of the inner cannula exceeded the length of the trach leaving a sharp edge to come in contact with the patients trachea.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4).As a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.One picture was attached.During the analysis conducted, it could be observed the back side of the packaging product; however, there is not evidence regarding to the failure mode reported.Both cannulas received were inserted into tracheostomy tube and it was observed that cannula overpass the length tube; thus, the failure mode reported is confirmed.The root cause of the reported issue was able to be determined as supplied item.Actions were taken to mitigate the reported issue: supplier started a project to replace the existing 2 cavity.No problems or issues were identified during this device history record review.
|
| |
|