H4: the subject device was manufactured on december, 2020 based on the provided 3 digit lot information.Therefore, 01dec2020 was selected.The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The device evaluation was unable to find any issues connecting the (blue) plastic connectors or connecting tube devices onto the a1 and a2 ports of the oer-elite reprocessor.In addition, the connecting tube (blue) plastic connectors were connected and disconnected multiple times on the a1/a2 ports of the oer-elite reprocessor and there was no signs of the lock levers/metal clips and (blue) plastic connectors coming off/popping off/coming apart.The click sound was verified and secured when connecting the plastic connectors to the a1 and a2 ports as well.The following findings were observed, however are considered non-critical and therefore do not present a potential for harm: there were scratches noted on both plastic connectors, scratches and stains on one of the plastic tubes and the other tube had scratches outside.Scratches on the metal connector were also observed.The device history record was unable to be reviewed for this device since the full lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, there was no abnormality with the connecting tube.Therefore, it is likely that the tube was not pushed enough (until it clicked) during connection or foreign material or the like got caught between the connecting part, which made the tube unable to be connected.The user can detect the event by conducting inspection as follows per the instructions for use (ifu): "9.Preparation and inspection: move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.5.6 inspecting the connectors connect each connector firmly by pushing until the connector clicks into place.After connection, pull the connector gently to confirm that it cannot be disconnected easily." olympus will continue to monitor field performance for this device.
|
|
The customer reported to olympus that the connecting tube cannot be connected to the channel connector in the reprocessing basin.The customer stated the connectors are falling apart.The connectors became detached too easily and at times, the connector hook becomes detached from the connector.There was a metal clip in the clasps that falls off, causing the plastic ¿wings¿ to come off the connector.There was no patient harm associated with the event.The medical device report (mdr) is related to the following patient identifiers: (b)(6)(model number: maj-2330) (b)(6) (model number: maj-2330) (b)(6) (model number: maj-2330) (b)(6)(model number: maj-2330) (b)(6) (model number: maj-2118) (b)(6) (model number: maj-2118): this report (b)(6) (model number: maj-2118) (b)(6) (model number: maj-2118) (b)(6) (model number: maj-2115) (b)(6) (model number: maj-2115) (b)(6) (model number: maj-2115) (b)(6) (model number: maj-2115) (b)(6) (model number: maj-2114) (b)(6) (model number: maj-2114) (b)(6) (model number: maj-2114) (b)(6) (model number: maj-2114) (b)(6) (model number: maj-2116) (b)(6) (model number: maj-2116) (b)(6) (model number: maj-2116) (b)(6) (model number: maj-2116) (b)(6) (model number: maj-2138) (b)(6) (model number: maj-2138) (b)(6) (model number: maj-2138) (b)(6) (model number: maj-2138).
|