Model Number IPN917285
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
It was reported that during morning quality check, the pump was unplugged to assess battery performance.When the pump was unplugged, it shut off.The user denies any battery alarms and states that the pumps are plugged in continuously.No patient involvement.As a result, "replaced power supply, they had display issues.Replaced cpm.Found umbilical cable heath had pulled away from connection, replaced cable".
 
Event or Problem Description
It was reported that during morning quality check, the pump was unplugged to assess battery performance.When the pump was unplugged, it shut off.The user denies any battery alarms and states that the pumps are plugged in continuously.No patient involvement.As a result, "replaced power supply, they had display issues.Replaced cpm.Found umbilical cable heath had pulled away from connection, replaced cable".
 
Additional Manufacturer Narrative
(b)(4).Returned for investigation was a martek power supply.Visual inspection of the power supply was performed and no abnormality was noted.The power supply was installed into a known good ac3 for functional testing.The system powered-up successfully.The power supply voltage check was performed per the ac3 series service manual and all output voltages were within specifications.The battery load test was performed.The iabp was run on battery until the iabp shut down.The battery was then left to charge in the iabp for over 9 hours.The full charge of battery was 13.0 volts.The iabp gave a proper alarm when disconnected from the ac power.Pumping was initiated.The unit ran for over 90 minutes along with the proper alarms "less than 20, 10, and 5 minutes remaining".A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of "shuts off when unplugged" is not confirmed.The returned power supply passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key17049720
Report Number3010532612-2023-00303
Device Sequence Number12510985
Product Code DSP
UDI-Device Identifier10801902172065
UDI-Public10801902172065
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K162820
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN917285
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2023
Supplement Date Received by Manufacturer11/28/2023
Initial Report FDA Received Date06/02/2023
Supplement Report FDA Received Date11/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-