| Model Number |
IPN917285 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during morning quality check, the pump was unplugged to assess battery performance.When the pump was unplugged, it shut off.The user denies any battery alarms and states that the pumps are plugged in continuously.No patient involvement.As a result, "replaced power supply, they had display issues.Replaced cpm.Found umbilical cable heath had pulled away from connection, replaced cable".
|
| |
|
Event or Problem Description
|
|
It was reported that during morning quality check, the pump was unplugged to assess battery performance.When the pump was unplugged, it shut off.The user denies any battery alarms and states that the pumps are plugged in continuously.No patient involvement.As a result, "replaced power supply, they had display issues.Replaced cpm.Found umbilical cable heath had pulled away from connection, replaced cable".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned for investigation was a martek power supply.Visual inspection of the power supply was performed and no abnormality was noted.The power supply was installed into a known good ac3 for functional testing.The system powered-up successfully.The power supply voltage check was performed per the ac3 series service manual and all output voltages were within specifications.The battery load test was performed.The iabp was run on battery until the iabp shut down.The battery was then left to charge in the iabp for over 9 hours.The full charge of battery was 13.0 volts.The iabp gave a proper alarm when disconnected from the ac power.Pumping was initiated.The unit ran for over 90 minutes along with the proper alarms "less than 20, 10, and 5 minutes remaining".A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of "shuts off when unplugged" is not confirmed.The returned power supply passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
|
| |
|
Search Alerts/Recalls
|